Environmental requirements, Ce mark, Specifications for sm semi micro balances – Scientech 10K Series User Manual
Page 22

Environmental Requirements:
This product is intended for indoor use at altitudes up to 2000 meters, Pollution Degree 1 or 2 in accordance with IEC 664 and transient overvoltages according to Installation
Categories (Overvoltage Categories) I, II and III. For mains supply, the minimum and normal category is II.
CE Mark:
All Scientech Electronic Balances carry the CE Mark and conform to the requirements of the following standards:
EN 55011: 1998
Limits and Methods of Measurement of Radio Disturbance Characteristics of Industrial, Scientific and Medical (ISM) Radio-frequency Equipment
EN 61326-1: 1997
EMC Requirements for Electrical Equipment for Measurement, Control and Laboratory Use
EN 61010-1: 1993
Safety Requirements for Electrical Equipment for Measurement, Control and Laboratory Use
Specifications for SM Semi Micro Balances:
Model
SM 50
SM 99
SM 124D
SM 128D
Weighing Mode
Single Range
Dual Range
(Floating Tare™)
Capacity
50g
99g
40g
120g
80g
120g
Readability
.00001g
.00001g
.0001g
Tare Range
0-80g
0-99g
0-120g
0-120g
0-120g
0-120g
Repeatability
(Standard Deviation)
.00002g
.00004g
.00002g
.0001g
.00003g
.0001g
Linearity
±.00003g
±.00003g
±.00003g
±.0002g
±.00003g
±.0002g
Stabilization Time
~8 Seconds
Weighing Speed
Adaptive
Display Update Interval
800ms
Data Interface
RS-232 bidirectional interface is standard. Baud rates are 300, 600, 1200, 2400, 4800, 9600 and 19,200.
Pan Size
Circular 3.5 inch diameter
Operating Temperature
5°C to 40°C
Relative Humidity
80 % for temperatures up to 31°C decreasing linearly to 50 % at 40°C
Power Supply
115 Volts/60 Hz ± 10 %
230 Volts/50Hz or 60 Hz ± 10 %
Windscreen
7.5W x 8.25D x 9.5H
Calibration Weight
50g
100g 100g
100g
10g
50g
40g
50g
Housing (WxDxH)
7.5" x 11.25" x 3.25"
Weight
~ 13.8 lbs
22