4. Set the temperature controller at the required temperature using the up and down arrows on
the keypad. See Appendix A for more details.
5. Final adjustment of the temperature controller may be required after the chamber has
reached operating temperature and this should be checked with a suitable thermometer
located in the chamber (A thermometer is not supplied with the chamber).
6. Where there are excessive amounts of fumes which may ignite when the door is opened
turn the fridge/freezer off before opening the door and then allow at least 30 minutes and
ventilate the room for the fumes to dissipate before switching the fridge/freezer back on. If
you are unsure leave off until the room is checked.
Temperature Controls
See Appendix A
Maintenance
The cabinet is finished with painted epoxy and to maintain appearance, should be wiped over with a
non abrasive household cleanser. The chamber is manufactured from formed plastic and may be
cleaned with a gentle cleaner. The door is also plastic lined and should only be cleaned with general
purpose cleaners.
As there is no internal drainage (so any fumes do not come into contact with the external
compressor motor) the catchment trough inside the fridge at the base will need to be
wiped out weekly to remove any condensate in the trough.
Trouble Shooting
SYMPTON
REMEDY
No Power
(Indicator Light is off)
1. Check fridge/freezer is plugged in and
power switched on.
2. Ensure mains power supply point is
functioning by using a test appliance on
power socket.
3. Check internal RCD has not tripped.
Failure to maintain temperature
(Indicator light is on)
1. Ensure the temperature controller set
point is above ambient.
2. Check the safety controller fitted is above
the main controller setting.
If the fault cannot be found call your distributor or the manufacturer quoting the serial number of the
unit from the manufacturers label.
Declaration of Conformity
Each product is thoroughly inspected and tested to not only ensure that it meets the specifications
provided, but to also meet Australian Electrical Standard AS3820 and EMC Standard
AS/NZ1044:1995,and therefore being accredited with a C Tick label.
Laboratory Equipment Pty Ltd – Spark Proof Fridges/Freezers Instruction Manual Page
2