LABEC AA20 User Manual
Page 6

Maintenance
Check the Safety Valve monthly by pulling the safety spring for 1 second to release steam whilst
under pressure. Wear Full Face Mask, Gloves and full length clothing and other PPE as required.
Ensure no product is inside the Autoclave during testing.
Clean drain and flush with boiling clean water regularly to prevent build up.
Ensure Pressure Gauge reads 105 KPA at 121ºc when in use. If not switch off immediately and have
the unit serviced.
Have the unit inspected by a registered boiler inspector annually. ( Ensure chamber is registered
with the appropriate body for pressure vessels)
The cabinet is finished with stainless steel and to maintain appearance should be wiped over with a non
abrasive cleanser. The chamber is manufactured from stainless steel and may be cleaned with a solvent
and if grained stainless steel chamber a stainless steel scratch pad may be used. The pad should be
rubbed in the same direction as the grain pattern of the stainless steel.
The silicone gasket should be cleaned with detergent only, ensuring it is dried completely after washing. To
prolong life it may be stored at cool temperatures in a container of water. Gasket lubricant is available to
provide prolonged life and better sealing on the gasket.
The locking pins should be regularly greased with high temperature grease to reduce binding in the guides.
WARNING: THIS AUTOCLAVE SHOULD NEVER BE CONNECTED TO MAINS WATER SUPPLY
Troubleshooting
SYMPTOM
REMEDY
No Power
(Indicator Light is off)
1. Check autoclave is plugged in and
power switched on.
2. Ensure mains power supply point is
functioning by using a test appliance
on power socket.
3. Check internal RCD has not tripped.
Failure to heat or maintain temperature
(Indicator light is on)
1. The temperature controller is on.
2. Handle ‘A’ is pulled forward activating
the microswitch.
3. The water level is above the low
water sensor and the low water
indicating lamp is off.
The timer does not start
1. Check the AL4 value is equal to set
value as this value activates the
timer.
5 Laboratory Equipment Pty Ltd Instruction Manual Autoclave AA13,16,18,20 Ver 2.51 2013