beautypg.com

Buffalo Filter BILF100 User Manual

Bilf100

background image

BILF100

Electrosurgical and Laser In-Line
Smoke Filter

INSTRUCTIONS

The Buffalo Filter Electrosurgical and Laser In-Line Smoke Filter

is used to filter smoke

plume, aerosols and debris generated during laser and electrosurgical procedures. Per NIOSH publication

no. 96-128, in-line filter protection is required when room suction is used for smoke plume capture.

Replacement filter for Stackhouse ILF100 Filter

Instructions:

1. Remove hose barb cap. Install ¼ inch tubing to the hose barb at the “suction” end of the filter.

Connect this tubing to wall suction.

2. Remove hose barb cap. Install ¼ inch tubing to the hose barb at the “patient” end of the filter.

Connect this tubing to the surgical laser, electrosurgical pencil adaptor or other appropriate surgical

instrument.

3. Install a fluid collection device (Buffalo Filter part number SCC1001) between the filter and the

surgical instrument if excessive fluid is expected. Failure to install a fluid collection device could

cause filter blockage and/or failure.

4. This

filter is considered a single use filter and should be removed and discarded after each

surgical procedure. Used filters are considered biohazardous waste; handle according to

appropriate hospital protocol.

For ordering information:

Buffalo Filter

5900 Genesee Street

Lancaster, New York 14086

(716) 835-7000
(800) 343-2324

(716) 835-3414 FAX

900240REVB DFU BILF100.doc