Buffalo Filter BILF100 User Manual
Bilf100
BILF100
Electrosurgical and Laser In-Line
Smoke Filter
INSTRUCTIONS
The Buffalo Filter Electrosurgical and Laser In-Line Smoke Filter
is used to filter smoke
plume, aerosols and debris generated during laser and electrosurgical procedures. Per NIOSH publication
no. 96-128, in-line filter protection is required when room suction is used for smoke plume capture.
Replacement filter for Stackhouse ILF100 Filter
Instructions:
1. Remove hose barb cap. Install ¼ inch tubing to the hose barb at the “suction” end of the filter.
Connect this tubing to wall suction.
2. Remove hose barb cap. Install ¼ inch tubing to the hose barb at the “patient” end of the filter.
Connect this tubing to the surgical laser, electrosurgical pencil adaptor or other appropriate surgical
instrument.
3. Install a fluid collection device (Buffalo Filter part number SCC1001) between the filter and the
surgical instrument if excessive fluid is expected. Failure to install a fluid collection device could
cause filter blockage and/or failure.
4. This
filter is considered a single use filter and should be removed and discarded after each
surgical procedure. Used filters are considered biohazardous waste; handle according to
appropriate hospital protocol.