W to – Allersearch Portable Nebuliser User Manual
Page 2
![background image](/manuals/659499/2/background.png)
Check the following, if the problem still persists, please contact your supplier.
Cause of failure
Causes and rectify measures
Power on, no operation
1. Check if the battery voltage is sufficient.
When the device’s red indicator light is
on, the battery power is too low.
Please replace the batteries immediatel y.
2.
Discard the waste batteries responsibly.
Check if the batteries are installed
correctly? Remove the batteries if
the equipment is not likely to be
used for some time.
Cannot nebulise or not
enough mist output
1. Check if the medicine cup is filled with the appropriate
amount of medicine.
2. The mesh in medicine cup is blocked by remaining
medicine. Clean medicine cup as per
Cleaning Instructions.
3. The mesh surface (mesh output) is covered by water
or medication.Remove water or medication from
mesh surface.
4. Medicine cup is not installed properly. Release the
medicine cup from main body and push it back again.
Power Supply
1.5V X 2, AA Size Alkaline battery
Power Consumption
2W
Active current consumption
650 mA
Standby current consumption
0.35mA
Battery life
Continuous operation:
1 hour (min) for brand new 1.5V X 2, AA,
alkaline battery or Ni-MH r echargeable
batteries (optional).
Full battery indication
2.2V, green light
Weak battery indication
2.0V ~2.2V, orange light
Low battery indication
2.0V, red light
Operation environment
10~40°C, 30%~80% RH
Storage/ Transportation environment
-20~70°C, 85% RH
Nebulisation rate
0.2mL/min,~1.0mL/min (Max)
Particle size
Time AUTO-OFF
Power on, operate 20 min. Then auto-off
Nebuliser auto-off
No medication
Medication bottle volume
2~8 mL (Max)
Ultrasonic Oscillation Frequency
Reusable parts life
103 KHZ~123 KHZ ± 1K
Medicine cup, mask, mouthpiece: one year -
based on use once per day
The specifications may be changed without prior notice.
* The product has been certified to meet the Harmonized Standard EN60601-1
and EN60601-1-2
1. If operating for the first time, insert 2 AA batteries as shown.
Suggestion: Use alkaline batteries or Ni-MH rechargeable
batteries (optional).
•
Remove batteries if the unit is not in use for long periods.
•
Discard the waste batteries responsibly.
2. Make sure that all parts such as mouthpiece or mask and medicine
cup are clean before use.
3. Push the exit button to seperate medicine cup from main body.
4. Pour the medicine into the medicine cup and close the cover.
5. Reattach the medicine cup to the main body until it clicks into position.
6. Secure mask or mouthpiece and begin breathing normally through
the device.
•
When the power button is pressed and the device starts working,
the green light will display and the timer will commence countdown.
•
If the indicator turns orange, it shows the battery is weak. If the
device is activated and the power indicator turns red, it means the
batteries are exhausted. Replace batteries immediately.
•
If there is no medicine inside the medicine cup, the indicator
light will turn red and flash four times before the device
turns off automatically.
W TO
After each use, please clean the following parts:
1. Remove the medicine cup and mouthpiece or mask. Discard all remaining medicine in the cup.
2. Pour in some hot water [about 100°C] and close the cover completely. Avoid excessive water.
3. Press the power button and operate 1-2 minutes to discard all remaining medicine completely.
4. Discard remaining water, clean and air dry medicine cup.
•
Clean the surface of main body with cotton swabs or a piece
of fine cotton fabric.
•
Clean the electrode in the main body and
medicine cup.
•
Please do not use volatile liquids to clean
the electrode, e.g. Benzene, thinner or
solutions containing sodium hypochlorite.
•
Clean the medicine cup,mouthpiece and mask
with clean water and store in a clean, safe place.
Disinfection
Medicine cup can be disinfected by boiling in hot water.After 10 to 30 minutes of boiling,
shake off excessive water from the cup and let it dry on a clean lint-free [paper] towel. Main unit,
mask and mouthpiece cannot be boiled.
Additional instructions for mesh care
Leave the mesh cap in a white vinegar solution [1 part vinegar diluted with 3 parts of water) for
1/2-1 hour. In addition, the mesh cap may be kept overnight in a cup of distilled water, in which
a tablet of dental ['false teeth') cleaner has been added. After cleaning with vinegar or dental cleaner,
be sure to shake off excessive solution, thoroughly but carefully rinse the mesh with distilled water.
Shake off excessive water from the mesh cap and let it air-dry on a clean, lint free [paper) towel.
Should the unit nebulise poorly even after following this procedure, contact your supplier.
2. INSTRUCTIONS FOR USE
3. CLEANING & STORAGE INSTRUCTIONS
4. TROUBLE SHOOTING
5. SPECIFICATIONS
~5 micron (MMAD)