Ionode IJ pH V5.1 User Manual
Page 39

Ionode IJ Instruction Manual
39
Symptom Possible
Causes
Remedy
Not within
Tolerance
(±20mV)
Contaminated
Standards
Replace standards
Correlating to
the Wrong
Reference
Electrode
Calculate what the
standard would
read using a
saturated KCl
Ag/AgCl reference.
See 5.1 ORP
Theory
Platinum not
Clean
See 4.4.1 pH
Cleaning
Electrode
Polarised
Replace electrode
Displays
0mV for
all
Solutions
Electrical
Short
Check connector
Internal Short Replace electrode
Displays
off Scale
Noise
See 5.5
Troubleshooting
Noisy