beautypg.com

Vernier Water Depth Sampler User Manual

Page 4

background image

© 2006 The Science Source · P.O. Box 727 · Waldoboro, Maine 04572

·Tel. 1-800-299-5469 e-mail us: [email protected] visit our

web site: www.thesciencesource.com

O:\_VST PAPERWORK\SENSOR PAPERWORK\SENSOR BOOKLETS\WDS\150101.DOC

November 27, 2006

4

All plants and animals need phosphates to grow. If the phosphate level in the water is too

high, excessive plant growth is the result. The Phytoplankton will grow out of control until their
own death rate becomes so large that the decay process uses up the available oxygen in the water
and other organisms die from this lack of oxygen. Ideally, rivers and streams have phosphate
concentrations of 0.1 ppm.

Nitrates and Nitrites:

Nitrates also promote the growth of Phytoplankton and when the concentration is too

high, their growth is uncontrolled. The decay of the dying plankton uses up the available oxygen
causing other animals to die.

Nitrites can cause fish to become ill because it reacts directly with the hemoglobin in their

blood, destroying its ability to carry oxygen. Concentrations below 0.06 ppm appear to have little
affect on cold water fish. Since nitrites rapidly become nitrates, they must be considered when
determining the nitrate concentration.

Sample Data Sheet


Experimenter's Name ___________________________________________________________
Date _____________________ Site # _______________________________________
Time _____________________

Weather conditions _____________________________________________________________
Air Temp _________________ Water Temp _____________________

Dissolved Oxygen

_______________

pH Reading

_______________

5 Day BOD

_______________

Phosphate

_______________

Nitrate

_______________

Other

_______________




Trouble Shooting:

Important! Always inspect the connections at each end of the shock cord (latex tubing) before each

use. These connections are fastened with nylon cable ties (available at most electronics hobby shops) and
should be replaced if evidence of wear is observed.

The trigger pin is designed to fit snugly in the hole through the side of the trigger assembly. Once the

sampler is armed, this pin may resist being pulled out of this hole. After several uses, this pin will wear