Calibrating the Humidity Display
1. Test the humidity display settings by placing the unit next to
another unit that is known to be accurate or by using the
Boveda One-Step Calibration Kit for at least 24 hours without
disturbing the environment
*
.
2. Determine the amount of calibration needed by taking the
display value and subtracting the expected value
**
. For
example, if you are expecting 70% but your unit displays 71%,
the amount of calibration needed is 71-70 = +1%.
3. Press the Set button for 3 seconds to place the unit into
calibration mode.
4. In calibration mode, you will see the current adjustment setting
flashing on the display. If this is the first time you are
calibrating your unit, it will show 0%. If you have already
calibrated the unit previously, you will see the current
adjustment.
5. Press the Min/Max/Clear button to adjust the display based
on your calibration. Pressing the button once will change the
adjustment by 1%. The adjustment range is +/-6% in the
following sequence: 0, +1, +2, +3, +4, +5, +6, -6, -5, -4, -3, -
2, -1, 0.
6. When you have completed your adjustment, press the Set
button one more time and it will take the unit out of
calibration mode. If you leave the unit in calibration mode for
more than one minute without pressing any buttons, the unit
will automatically will change back to operations mode.
* We recommend the Bovida One-Step Calibration Kit over other
testing methods. In no case do we recommend using the salt test.
This test is not accurate and can damage and void the warranty on
your Caliber IV
tm
.
** The Caliber IV
tm
can be adjusted +/- 6%. If you find the unit
is more than 6% out of calibration, please call us so that we can
troubleshoot with you or have you send it back for a factory
calibration.
Restoring Factory Calibration
To restore your Caliber IV
tm
back to factory calibration, simply
follow the calibration instructions above and adjust the display
back to 0%. This will restore the unit back to factory calibration.
Limited Warranty
The Caliber IV
tm
is covered by a limited warranty against defects
from materials or workmanship for a full 2 Years from the date of
purchase. Any misuse or excessive wear and tear is not covered
under the limited warranty. Keep your receipt in a secure location
as proof of purchase.
Contact Information
If you have any questions or warranty inquiries please call us
directly at 877-627-4798 or email [email protected].
Caliber IV
tm
is a trademark of
Western Humidor Corporation, USA
www.westernhumidor.com
And distributed by Cigar Oasis, Inc.
79 Heisser Court
Farmingdale, NY 11735
www.cigaroasis.com