Nextteq Thermal Ring Kit User Manual
Thermal ring introduction
Results from under-reported or over-reported chemical levels can cre-
ate an unsafe condition. The Thermal Ring increases safety and reduc-
es risk by allowing you to accurately correct for temperature changes
on-the-spot, without additional equipment like alcohol and infrared
thermometers.
Because detector tubes are calibrated at a specific temperature, in-
creases or decreases from this temperature can impact the reading.
The instructions that come with each pack of detector tubes advise if
there is a need for temperature correction. (For a complete list of tubes,
and indication of those needing temperature correction, consult Gas-
tec’s Detector Tube List or the Gastec Handbook, both available from
Nextteq, LLC.) The instructions will provide either a table with correc-
tion factors or a table listing true concentrations based on the ambient
temperature. For accurate results, ensure the tube has been at ambient
temperature for 15 minutes before testing.
The Thermal Ring, which is af-
fixed to all new Gastec GV-100
pumps, is perfectly calibrated
to work with the Gastec Gas
Sampling Pump and Detec-
tor Tube Systems. Please see
other side for instructions on
affixing the Thermal Ring and
warning label to a GV-100
pump in the field.
Thermal Ring Introduction
Thermal Ring Instructions
NEXTTEQ, LLC
8406 Benjamin Rd. Suite J Tampa, FL 33634
Tel: 813-249-5888 Toll Free Tel: 877-312-2333
Fax: 813-249-0188 Toll Free Fax: 877-312-2444
Email: [email protected] Web: www.nextteq.com
© Copyright 2005 Nextteq, LLC
Nextteq is Gastec’s exclusive U.S master wholesale distributor. Gastec
tubes and pumps are manufactured by the Gastec Corporation. Gastec
is a registered trademark of the Gastec Corporation. Nextteq is a regis-
tered trademark of Nextteq, LLC.
P/N INS-LAB-GV-100 Rev A 7/05