Power Plate pro5HP User Manual
Page 8
against, and have passed the applicable
requirements of relevant electromedical
standards, including EN 60601-1-2:2007.
The pro5™ and pro5 AIRdaptive™ models
must emit electromagnetic energy in
order to perform their intended function.
Nearby electronic equipment may be
affected by this emission. Similarly
the pro5™ and pro5 AIRdaptive™
models may themselves be affected by
electromagnetic emissions from other
equipment in the vicinity.
The pro5™ and pro5 AIRdaptive™
models are intended for use in the
electromagnetic environment specified
below. The customer or user of the
pro5™ or pro5 AIRdaptive™ model should
assure that they are used in such an
environment:
– Floors should be wood, concrete or
ceramic tile. If the floor is covered with
synthetic material, the relative humidity
in the room should be at least 30%.
– Mains power quality should be that
of a typical domestic, commercial or
hospital environment as appropriate.
– The pro5™ and pro5 AIRdaptive™
models should not be used during
power mains interruptions
– Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical commercial
or hospital environment.
The pro5™ and pro5 AIRdaptive™
models are intended for use in an
electronic environment in which radiated
radio-frequency (RF) disturbances
are controlled.
The customer or user of the pro5™
or pro5 AIRdaptive™ model can help
prevent electromagnetic interference by
maintaining a minimum distance between
portable and mobile RF communications
equipment (transmitters) and the
pro5™ or pro5 AIRdaptive™ model
as recommended below, according
to the maximum output power of the
communications equipment:
The adequate separation distance
to provide base immunity to RF
disturbances is 1.0m.
Retain this user manual for your future
reference.
•
•
•
•
08