Operation, Overloads, Safety – Dillon CT Handwheel Stand User Manual
Page 2: Maintenance

for attaching an inline tension grip. If it does not, alignment may be achieved with a
tension plate accessory or by using shims (not included) between the force gauge and
mounting plate. All five holes are M6 x 1.0mm. Dillon offers flexible coupler/adapters
if thread conversion is required.
Operation
Turn handle clockwise to raise the crosshead and counterclockwise to lower it.
The CT Test Stand is capable of overloading force gauges or itself. Keep a watchful
eye on the force gauge to insure forces remain within limits of the gauge and stand.
CAUTION: Test stand or gauge damage from overloads is not warranty covered.
Do not run the crosshead arm or mounting plate into the top or bottom of the stand.
Applying torque once contact occurs may result in overload damage to the stand.
Overloads to the test stand may result in non-smooth operation, particularly in specific
crosshead locations and/or handle positions. Overload of this type calls for stand
replacement.
Safety
WARNING: Pinching hazards exist with moving parts. Keep hands clear of all
moving parts other than the handwheel. Some samples release a large amount of energy
when they fail or separate. When working with such samples, insure some protection
exists for the operator such as a safety shield or part restraint.
Maintenance
Remove particle build up by rubbing the guide rods and acme threads with a
disposable cloth. The bushings are oil impregnated for long use. The guide rods and
acme threads may be coated with light general purpose oil if desired.
If the stand becomes difficult to turn under no load, loosen screws at the top and
bottom of the guide rods. Move crosshead to its bottommost position and tighten the lower
screws. Move the crosshead to its topmost position and tighten the upper screws.
Check for smoothness through the travel range. If still binding, stand may have been
overloaded. Seek replacement.
an Avery Weigh-Tronix Brand
USA office
1000 Armstrong Drive
Fairmont, Minnesota 56031 USA
Sales +01 (507) 238-8796
Service +01 (507) 238-4461
UK office
P. O. Box 9533
Smethwick
West Midlands B66 2TE U.K.
Sales +44 870 444 6132
Service +44 870 901 4116
11/06