Error Codes
Error codes are shown in the main display for approx. 2 seconds, after which the program automatically returns to the
weighing mode.
Display
Cause
Solution
No segments appear on
No AC power is available
Check the AC power supply
the display
The AC adapter is not
Plug in the AC adapter
plugged in
Battery is dead
Replace the battery
Recharge the battery using
an external recharging device
h
The load exceeds the
Unload the scale
scale capacity
l
The load plate is not in place
Place the load plate on the
scale
Something is touching
Move the object that is
the load plate
touching the load place
e 01
Display capacity exceeded:
Decrease the weight on
Value to be output cannot be
the scale
shown on the display
e 02
Calibration parameter
Calibrate only when zero
not met; e.g.:
is displayed
– scale not zeroed
Press t to tare the
scale
– scale is loaded
Unload the scale
e 09
When gross - zero, no tare
Tare the scale
e 10
The t key is blocked
Press [CAL/CF] > 2 sec. to clear
when there is data in
the tare memory and release the
the second tare memory
tare key
(net-total) – only 1 tare function
can be used at a time
e 11
Value input is not allowed for
Press t
second tare memory
e 22
Weight is too light or there is
Increase the reference quantity
no sample on the balance
e 30
Interface port for printer
Contact your local
output is blocked
Service Center
Max. weighing range is less
The scale was
Place the weighing pan on
than indicated under
turned on without the weighing
on the scale
”Specifications”
pan in place
and press [ON/OFF] to turn on the
scale
The weight readout
The scale has not
Calibrate/adjust the scale
is obviously wrong
been calibrated/adjusted
The scale was not
Tare before weighing
tared before weighing
If any other errors occur, contact your local Service Center!