8

Error Codes

Error codes are shown in the main display for approx. 2 seconds, after which the program automatically returns to the

weighing mode.

Display

Cause

Solution

No segments appear on

No AC power is available

Check the AC power supply

the display

The AC adapter is not

Plug in the AC adapter

plugged in

Battery is dead

Replace the battery

Recharge the battery using

an external recharging device

h

The load exceeds the

Unload the scale

scale capacity

l

The load plate is not in place

Place the load plate on the

scale

Something is touching

Move the object that is

the load plate

touching the load place

e 01

Display capacity exceeded:

Decrease the weight on

Value to be output cannot be

the scale

shown on the display

e 02

Calibration parameter

Calibrate only when zero

not met; e.g.:

is displayed

– scale not zeroed

Press t to tare the

scale

– scale is loaded

Unload the scale

e 09

When gross - zero, no tare

Tare the scale

e 10

The t key is blocked

Press [CAL/CF] > 2 sec. to clear

when there is data in

the tare memory and release the

the second tare memory

tare key

(net-total) – only 1 tare function

can be used at a time

e 11

Value input is not allowed for

Press t

second tare memory

e 22

Weight is too light or there is

Increase the reference quantity

no sample on the balance

e 30

Interface port for printer

Contact your local

output is blocked

Service Center

Max. weighing range is less

The scale was

Place the weighing pan on

than indicated under

turned on without the weighing

on the scale

”Specifications”

pan in place

and press [ON/OFF] to turn on the

scale

The weight readout

The scale has not

Calibrate/adjust the scale

is obviously wrong

been calibrated/adjusted

The scale was not

Tare before weighing

tared before weighing

If any other errors occur, contact your local Service Center!