2.5 – Common Issues

The Picker software does not connect to the instrument or show an “Online” status.

- Ensure the picker is on before the software is activated.

- Ensure the picker communication cable is connected to COM 1 of the computer.

It cannot be connected to any other COM port.

- Ensure that wish84.exe is NOT running prior to starting the picker software, and

ensure that no other picker software windows are running. This can be verified
within Windows Task Manager.

- Ensure window firewall is not blocking wish84.exe.


The picker is barely missing colonies by a few mm.

- Ensure the camera cap is completely off, and the camera is connected.

- Check the focus on a colony plate by opening the Interactive Picker Imaging

program. Check the CP7200 manual for details.

- Check all pins for bends and ensure that they are installed properly.

- Go through the CP7200 calibration procedure. Use the white mounting block

whose height is closest to agar levels of source plates while in the source nest.
NOTE: Failure to use a similar height block will cause all pins to be off by a very
small distance.

- Do a test pick after a successful calibration procedure. If colonies are still off,

contact Hudson Robotics for advice.