beautypg.com

Gascoseeker 2-500 operation, Sample start, Switch on – Gas Measurement Instruments Gascoseeker 2-500 User Manual

Page 2: Sample finish, Switch off, Zero mode, Press measure button, All lcd segments are displayed, Switch off is automatic after 30 minutes running

background image

Sample Start

Press MEASURE Button
once to switch on pump.
‘SAMPLE’ flag indicates that
pump is running and display
shows analysis of sample:

GASCOSEEKER 2-500

OPERATION

once to switch instrument ON. Instrument performs self-check and start-up
sequence. Pump also operates for a short period during this sequence.

The following display indicates instrument
version identifier and software version
number:

A bar graph indicates battery capacity and
background illumination is activated:

After a few seconds the instrument is
ready for use, indicated by a flashing
digital display:

BUTTON 1

BUTTON 2

OFF

MEASURE

PUMP ON
PUMP OFF

ZERO

GASCOSEEKER 2-500

Switch ON

Press MEASURE Button

%
V
SAMPLE STORE

ZERO

FAULT

CHECK
BAT

10000PPM

8000

6000

4000

2000

0

BAT

100

80

60

40

20

0

All LCD segments are displayed:

%

100

80

60

40

20

0

Display flashing

OFF

MEASURE

PUMP ON
PUMP OFF

%

SAMPLE

100

80

60

40

20

0

Sample Finish

A single press of the MEASURE button

, when the pump is running, stops

the pump and stops sampling. This is indicated by the absence of both ‘SAMPLE’
flag and flashing display.

Switch OFF

Press the MEASURE button
twice in rapid succession to
switch the instrument OFF.
‘OFF’ appears in the display and
the instrument will shut down.
(To cancel this action, press the
MEASURE button once).

OFF

MEASURE

PUMP ON
PUMP OFF

%

100

80

60

40

20

0

Switch off is automatic after 30 minutes running.

Zero Mode

Make sure that the instrument is switched OFF and in FRESH AIR before
zeroing.

Press the ZERO button
once to enter zero mode.
The instrument will enter zero
mode with similar test sequence
to measure mode except for
‘ZERO’ flag, indicated in the
status display as shown.

OFF

MEASURE

ZERO

GASCOSEEKER 2-500

%

100

80

60

40

20

0

ZERO

The instrument is then zeroed automatically.
If this operation does not reset the zero and remove the ‘ZERO FAULT’ or
‘CHECK ZERO’ flags, the instrument should be returned to the service centre
for checking.
The instrument switches OFF automatically from the ZERO mode.
The instrument can only be zeroed by switching OFF from the MEASURE mode.

V