4. Place the consolidometer with smaple on the load platform and ad-

just the top cross arm adjusting screw until it makes contact with the

load pad in the consolidometer. Position the dial indicator over the

cross arm screw until it makes contact with the load pad in the con-

solidometer. Position the dial indicator over the cross arm screw and

adjust the indicaotrl to the desired setting. the seating load can now

be applied by backing off on the load holding screw, making sure

there is sufficient clearance between the screw and the lever arm.

5. Refer to ASTM D2435, D4546, AASHTO T216 OR BS 1377:PART

5:1990 for additional information on specific test procedures.

6. ASTM D-2435 requires that corrections for vertical deformations in

the frame must be made by sing a steel calibration disc the same

height as the sample and 0.04” smaller in diameter.

Dead Weight Consolidation Load Chart

2.5” Dia. Sample, Area = 4.9087 in

2

Weights Added 10:1 Beam = Sample Load

Quantity

Lb Wt.

Total lbs

TSF

1

.852

8.52

1/8

1

.852

17.04

1/4

1

1.704

34.09

1/2

1

3.409

68.18

1

1

6.818

136.35

2

1

13.635

272.7

4

1

27.27

545.4

8

2

27.27

1090.8

16

4

27.27

2181.6

32

Weights Sets

Kg Weight Set

1Kg

4 Kg

8Kg

HM-1122

4

3

2

HM-1123

4

5

5

16 tsf Weight Set

1/8 tsf

1/4 tsf

1/2 tsf

1/8 tsf

2 tsf

4 tsf

HM-1120

2

1

21

1

1

1

32 tsf Weight Set

HM-1121

Add (4) 4 tsf weights to the 16 tsf set listed above