Calibration

• Obtain a force-measuring gauge appropriate for measuring compressive

force from 20 to 200 lbf and acceptable as reference standard.

• If the measurement standard will not fit under the Penetrometer’s needle

adapter, remove the breather plug, H-4133.8.4 from the cylinder, and the

needle adapter, H-4133.8.9, from the piston stem.

• Remove the gauge & cylinder assembly, H-4133.7A, by taking off the lock

nut, H-4133.8.5.

• Install the cylinder assembly into the measurement standard test frame and

secure with the lock nut and check in 20 lbf steps from 20 to 200 lbf.

• Apply force and check in 20 lbf steps from 20 to 200lbf.
• If the readings depart from the standard more the ± 2 lbf return the gauge

& cylinder assembly to the factory for service.

Maintenance

Keep the units clean at all times. If re-lubrication is necessary, use Teflon based

grease to avoid particulate accumulations.