beautypg.com

Humboldt H-4204 Proving Ring Penetrometer User Manual

Page 4

background image

Rezeroing the Instrument
When reading has changed at any time or if any part is changed in the

penetrometer, the instrument should again be zeroed as described in the

Zeroing Instrument section above.
Operation

Be certain that the dial indicator has been set to zero position. Select the site to

be tested and clear the test location so that a flat and clean surface is available

for testing.
Hold the assembly vertically on the test location. Grasping the handle firmly,

push the cone point down into the soil at a steady uniform rate until the top of

the cone goes just below the surface.
Determine the dial indicator reading as follows:
Each subdivision has a value of one. The small dial reads from 0 to 10 with one

division being equal to one full rotation of the large dial. Each value on the small

dial has a value of 100. Thus, one complete rotation of the large dial is 100 and

is recorded as 1 on the small dial.
To determine the penetration resistance (bearing capacity) in pounds per square

inch, record the dial reading of both the inner and outer dials. Use this number

in refering to the Proving Ring Calibration Chart provided to determine the pen-

etration resistance (bearing capacity) in pounds force at the depth being tested.
In order to measure the penetration resistance at 6", 12" or other depths, push

the cone point down into the soil until the 6", 12" or other marks on the rod

just hit the surface and record the dial indicator reading. Refer to the calibration

table to read the pounds force value.
Note: An approximate resistance (bearing capacity) is obtained by dividing the

penetration load by the cone base area. The cone base area is actually 0.983

square inch. But in order to make correction for the cone soil friction, the base

area should be taken as 1 square inch. Hence the penetration load reading

in pounds is numerically equal to the penetration resistance or cone index

in pounds per square inch.
Return the dial indicator reading to zero position by pressing the release

button on the pointer stem housing. Support the release button from behind

while pressing.
In soils of very low resistance, it maybe desirable to utilize a deeper penetration.

The 6" marks or other marks, which the user may wish to scribe on the

extension, may be used as the stopping point rather than the top of the cone.

These penetration depths allow correlation with laboratory tests as well as the

standard penetration depth.