beautypg.com

Humboldt H-4040 Series Water Level Indicators User Manual

Page 2

background image

GENERAL
The H-4040 Series Water Level Indicators are self-contained instruments for

determining water levels in wells, bore holes, etc. Each unit consists of a hand-

held wire spool with indicator module, measuring cable and probe. The spool,

constructed ABS plastic and steel is equipped with a brake and clip for probe

storage.
The indicator module has both audible and visual indicators to show when

water has been detected. A standard 9-volt battery is the only power

source required. The multi-conductor cable does not require any additional

connections or ground wires. The wire used is high-strength, hard-drawn,

copper-coated steel for stretch resistance. Markings are permanently

embossed on the wire.
The probe is constructed of stainless steel and Teflon* for environmentally

sensitive testing.
FEATURES OF MODULE
The following features are included on the front panel of the

CONTROL MODULE:
• The POWER SWITCH controls power to the instrument and is labeled OFF.

The unit in ON when the switch is in the upper-most position.

• The TEST SWITCH, labeled TEST, is used to check the battery condition.
• The RED LIGHT at the top of the panel serves two functions. First, it

indicates the condition of the battery. It also works in conjunction with the

horn, as an indicator to announce when water has been located. The light is

not labeled.

• The HORN, in the center of the panel, provides audible signal when water is

contacted. The horn is not labeled.

• The SENSITIVITY CONTROL provides a means to adjust the instrument and

compensates for varying water conditions. See Section 3.0 for details.

OPERATION
IMPORTANT! Set the SENSITIVITY CONTROL mid-range or higher (toward the

+ sign), anytime you are searching for water of an unknown conductivity. To test

battery, turn power on by sliding the left switch up toward the top of the

panel. Press right switch labeled TEST (red light will glow brightly if battery is

fully charged). See Section 7.0 for important details regarding battery condition.
To test the system, the power must be on and the SENSITIVITY CONTROL set

mid-range or higher (toward + sign). Dip the probe in water and the light and

buzzer should function. You can also wet the tip of your thumb and pointer

finger with water or saliva and simultaneously touch the two needles at the

end of the probe to get the same effect.
For normal operation, simply lower the probe down the well or bore hole.

When the probe touches water, the buzzer and light will operate. When the

probe is lifted clear of the water, the indicators will shut off. The SENSITIVITY