CF 27
5
4/2001
Bridge Balance
The Bridge Balance pot provides up to 68% of full-scale tare weight adjustment when using a 350-ohm, 3mV/V load
cell. The DCM 437 can then be adjusted to provide full span output with only about 10% of the cell’s range.
DCM 437 Gain Switch
Bridge Output Signal for:
Gain Switch
Position “ON”:
0 to 20 mA
4 to 20 mA
1
50mV to 62.5mV
40mV to 50mV
2
25mV to 50 mV
20mV to 40mV
3
12.5mV to 25mV
10mV to 20mV
4
6.25mV to 12.5mV
5mV to 10mV
5
3.125mV to 6.25mV
2.5mV to 5mV