Section 7
Troubleshooting Guide
Problem
Possible Cause
Possible Solution
Low protein
Low levels of protein
Increase load on column
concentration
in sample
Concentrate sample in
SpeedVac
High protein
High levels of protein
Decrease the load
concentration
in sample
applied to column
Buffer does not flow
Frit is clogged with
Prefilter sample prior
through column
sample debris
to loading
Protein does not bind
Incorrect column chosen
CEX column is for
to column
basic proteins; AEX
column is for acidic
proteins
Ionic detergents can
Avoid using detergents
prevent sample from
with opposing net
binding
charges to that of ion
exchange column.
Detergents can be
removed using
Bio-Beads
®
SM-2
macroporous beads
(catalog #152-3920)
Ionic strength too high
Dilute sample or desalt
pH not correct
Check pH of loading
buffer
Protein does not elute
Salt concentration too
Increase salt
from column
low in elution buffer
concentration to 2 M
and/or change pH
Sample is dilute after
Excess packing buffer
Remove excess packing
centrifugation
was not removed
buffer by centrifugation
prior to loading sample