NOTE:
Amico recommends that you have a blank space
between outlets and slides.
Maximum 3 outlets may be ganged together
without additional support.

5-1/2”

[140]

3”

[76]

5-27/32”

[148]

5”

[127]

5-3/4”

[146]

1-1/4”

[32]

3/8”

[10]

1-7/8”

[48]

MEDVAC

Ceiling Outlets

DISS, Ohmeda compatible, Chemetron compatible, Puritan-Bennett and Oxequip/MedStar compatible

When wall outlets are ganged
together, ensure that they
are 5” from center to center.
DO NOT let the notches on
the mounting plate pass the
adjacent holes on the next
mounting plate.

Do not bury the outlet during
the drywall and plaster appli-
cation. Ensure that the cover
over the outlet is in place un-
til the “Latch-valve assembly”
can be installed.

Inch

[mm]

Console Outlets

DISS, Ohmeda compatible, Chemetron compatible, Puritan-Bennett and Oxequip/MedStar compatible

TOP VIEW

1/2" [12.7 mm] O.D. (3/8” nominal) type "K"

copper pipe, rotates 360 deg.

for entry from any angle.