Preventative maintenance – Amico British Standard Outlet Slim User Manual
Page 7
www.amico.com
7
Regular inspection and maintenance of the Outlets will prolong its life and reduce the possibility of sudden, inconvenient
component failures. The use of damage probes or faulty equipment may require further maintenance.
Outlets should be subjected to regular inspection and testing as follows:
Monthly;
a. Visually inspect the Outlet for signs of damage.
b. Check to see if the Front Latch Assembly (2nd Fix) operates smoothly with the gas specific probe.
c. If there is any disagreement with the operation of the Front Latch Assembly (2nd Fix), remove the unit for closer
inspection/repair/replacement if necessary.
Annually;
a. Remove Trim Plate and clean debris from the mounting box.
b. Remove Front Latch Assembly (2nd fix) and inspect the barrel for any damage.
c. Inspect the Rough-in Assembly (1st fix) for any damage. If damage or leaking replace components.
d. Test Front Latch Assembly (2nd fix) for correct operation (connection and disconnection of the probe) using a
blank probe.
Preventative Maintenance
- Artwall (6 pages)
- Ceiling Column Manual Retractable (16 pages)
- Ceiling Column Motorized Retractable (20 pages)
- Ceiling Column Rigid (Stationary) (12 pages)
- Ceiling Column Rigid LITE (Stationary) (12 pages)
- Recessed Console Unit (8 pages)
- Freedom Series Recessed (12 pages)
- Freedom Series Surface Mounted (8 pages)
- LDR Headwall (8 pages)
- Majestic Series Double Tier Recessed Headwall (6 pages)
- Majestic Series Double Tier Surface Mounted Headwall (8 pages)
- Majestic Series Single Tier Surface Mounted Headwall (6 pages)
- Majestic Series Single Tier Surface Mounted Headwall (8 pages)
- Majestic Series Single Tier Recessed Headwall (8 pages)
- Noble Custom Cabinet (7 pages)
- Power Column (7 pages)
- Regal Series Recessed Vertical Flatwall (8 pages)
- Regal Series Surface Mounted Vertical Flatwall (7 pages)
- Security Recessed Console Box Assembly (12 pages)
- Surface Mount Console Unit (6 pages)
- Hydraulic Patient Transfer Stretcher (S-H-300) (20 pages)
- Hydraulic Patient Transfer Stretcher (S-H-300) (11 pages)
- AHM Monitor Arms (Adjustable Height) (34 pages)
- Barcode Dock Mount (15 pages)
- Condor Adjustable Height Channel (20 pages)
- Condor Lite Fixed Height Channel (31 pages)
- Condor Lite Fixed Height Channel (12 pages)
- Falcon Combo Arm (44 pages)
- Light-Duty Monitor Arm (20 pages)
- Monitor Roll Stands (12 pages)
- CSA Duplex RED Horizontal Tank Mount (36 pages)
- NFPA Duplex CCD Horizontal Tank Mount (51 pages)
- NFPA Duplex RED Horizontal Tank Mount (40 pages)
- 12" Floor Pedestal (8 pages)
- 2 in 1 Alarm System (56 pages)
- Alarm Valve Combo Unit (32 pages)
- Alert-3 LCD Alarm (28 pages)
- Automatic Cryogenic Liquid Solenoid Manifold (32 pages)
- Ball Valve (8 pages)
- British Standard Outlet (18 pages)
- Chemetron Compatible Ceiling Column Outlet (24 pages)
- Chemetron Retro-fit Gas Control Panel (4 pages)
- Dome Loaded Manifold CSA (44 pages)
- Dome Loaded Manifold Nema-4 NFPA (44 pages)