Emc-inspection report, Table 1: emission table 2: immunity (all devices) – Amico Gamma 50 LED User Manual
Page 15
www.amico.com
15
EMC-Inspection Report
Guidance and Manufacturer’s Declaration – Electromagnetic Emissions
The Gamma 50 LED is intended for use in the electromagnetic environment specified below. The customer or the user of the
Gamma 50 LED should assure that it is used in such an environment.
Emission test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The Gamma 50 LED uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are
not likely to cause any interference in nearby electronic
equipment.
RF emissions
CISPR 11
Class B
The Gamma 50 LED is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply network
that supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class C
Voltage fluctuations / flicker emissions
IEC 61000-3-3
Complies
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The Gamma 50 LED is intended for use in the electromagnetic environment specified below. The customer or the user of Gamma
50 LED should assure that it is used in such an environment.
Immunity test standard
IEC 60601 test level
Compliance level
Electromagnetic environment - guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
±6 kV contact
±8 kV contact
±8 kV air
±15 kV air
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the
relative humidity should be at least 30%
Electrical fast transient /
burst
IEC 61000-4-4
±2 kV for power
supply lines
±2 kV for power
supply lines
Mains power quality should be that of a typical
commercial or hospital environment.
Surge
IEC 61000-4-5
±1 kV line to line
±2 kV line to earth
±1 kV line to line
±2 kV line to earth
Mains power quality should be that of a typical
commercial or hospital environment.
Table 1: Emission
Table 2: Immunity (All Devices)