5.0

DIAGNOSTICS & REPAIR


PROBLEM

CAUSE

SOLUTION

No functions operate

No power to unit

Check power cord

Is it plugged into receptacle?

Green Power-on light

Electrical power cord not

does not glow

properly plugged into

Check cord for breaks.

control module

Power cord plugged

Check for power to receptacle

in but no functions

operate

Bad handset

Replace handset

Bad control module

Replace control module


6.0

CLEANING & MAINTENANCE


The chair is subject to wear and tear resulting from the place, length, and other conditions of use.

For longer use, periodically inspect all parts and have them replaced as needed.

check the cushions for rips or tears

check operation of the brake pedals

check set screws on brake pedals

check all bolts and nuts for tightness

Your maintenance department should inspect the chair periodically ( at least every 6 months) for

proper operation.

Every six months minimum:

lubricate all slides and pivot points with FDA approved light grease

check all electrical wiring for any frays, kinks and or deterioration

ensure that lock and steer castor operate properly


Any discrepancies noted during inspection must be corrected before using chair.

CLEANING

Any ?uid spilled on the seat, back or leg cushions should be removed as quickly as possible.

All painted metal surfaces should be wiped with a clean, soft cloth weekly.

All moving parts should be lubricated occasionally with light approved oils to help insure quiet,

dependable operation.

Before using any cleaning product, test it on the underside of the chair seat. Never use any vola

tile ?uid (thinner, gasoline, etc) that could cause discoloration or deterioration.

Use mild soap and water or a nonabrasive hospital grade detergent disinfectant to clean the chair.

First, wet a cloth with the cleaning product and wring it dry, then gently wipe off the chair.