The following table lists the error messages and their associated meanings:
Error
ASSOCIATED MEANING
TYPICAL CAUSE
1
Lamp failure
Lamp has too low an output for proper
turbidity measurement.
2
General Calibration failure:
There is not sufficient signal
between the 0.02 NTU and 10.0
NTU standards used for
calibration
Either the wrong standards were used in
calibration or there is an internal sensor
failure.
3
General Calibration failure:
There is not sufficient signal
between the 10.0 NTU and 100.0
NTU standards used for
calibration
Either the wrong standards were used in
calibration or there is an internal sensor
failure.
4
General Calibration failure:
There is not sufficient signal
between the 100.0 NTU and
1000.0 NTU standards used for
calibration
Either the wrong standards were used in
calibration or there is an internal sensor
failure.
Err
Multiple error conditions are met