Specifications – HF scientific Micro1000 Laboratory Turbidimeter - 0-10,000 NTU User Manual
Page 7

Rev. 3.2
1
1.
SPECIFICATIONS
MICRO 1000
MICRO 1000IR
Instrument Conformance
Conforms to specifications set forth in
EPA method 180.1 (Nephelometric
Method)
§
Conforms to specifications set forth in
ISO 7027: Water Quality –Determination
of Turbidity
Light Source
Tungsten Filament Lamp
Infrared LED
Measurement
Options and Ranges
NTU
FNU
EBC
NEPHELO
FAU
Ratio on or Ratio off
0 – 10,000
--------
0 - 2,450
0 – 67,000
--------
Ratio on or Ratio off
0 – 10,000
0 – 10,000
0 – 2,450
--------
0 – 10,000
Signal Averaging
Yes
Yes
Auto Ranging
Yes
Yes
Accuracy
†
0 – 1000 NTU
1000 – 4000 NTU
4000 – 10,000 NTU
± 2% of reading + 0.01 NTU
± 5% of Reading
± 10% of Reading
± 2% of reading + 0.01 NTU
± 5% of Reading
± 10% of Reading
Repeatability
± 1% of reading or ± 0.01 NTU
whichever is greater
± 1% of reading or ± 0.01 NTU
whichever is greater
Resolution
0.0001 NTU below 10 NTU
0.0001 NTU below 10 NTU
Response Time
Less than 6 seconds
Less than 6 seconds
Serial Input/Output
RS-232 Directional
RS-232 Directional
Certifications
CE, ETL (UL & CSA )
CE, ETL (UL & CSA)
Power Supply
15 V DC Wall Mount
15 V DC Wall Mount
Display
Backlit Multi-Line LCD
Backlit Multi-Line LCD
Miscellaneous
Specifications
1 - Built in Diagnostics
2 - Security Access Code
1- Built in Diagnostics
2- Security Access Code
Operating Temperature
Range
0° C – 50° C (32 ° F - 122° F)
0° C – 50° C (32 ° F - 122° F)
Sample Flow Rate
Through The Flow Cell
0.5 – 1.5 gpm (2-6 L/min)
60 psi max pressure (414 kPa)
0.5 – 1.5 gpm (2-6 L/min)
60 psi max pressure (414 kPa)
Air Purge
0.1 SCFM at 10 PSIG
0.1 SCFM at 10 PSIG
Shipping Dimensions
12” x 14” x 12”
(35.6cm x 30.5cm x 30.5cm)
12” x 14” x 12”
(35.6cm x 30.5cm x 30.5cm)
Shipping Weight
7 lb (3.18 kg)
7 lb (3.18 kg)
Warranty
1 - Year
1 - Year
§ The specifications found in EPA method 180.1 are essentially the same as the specifications set out in method
2130B of the Standard Methods for the Examination of Water and Wastewater 19th edition and the
specifications set out in ASTM Standard Method D1889-94. This turbidimeter meets, or exceeds, the
specifications set forth in these methods.
† Instrumental accuracy measured under controlled laboratory conditions at 25°C (77°F)