beautypg.com

Gossen MAVOSPEC Communication Protocol User Manual

Page 3

background image

GOSSEN Foto- und Lichtmesstechnik GmbH

3

Attribute name:

"integrationTime"

- sets integration time to be used by MAVOSPEC during measurement.

The min and max attributes contain the defined range of allowed values

for the integration time and cannot be changed.


"repeatCount"

- sets the number of measurement repetitions. The measurement result

send to the PC is the average over the given number of measurements.


"autoIntegration"

- turns on/off automatic mode - the effective integration time in this mode

will be calculated based on available amount of light. The calculated

integration time is stored in integrationTime.


"maximumIntegrationTime"

- defines maximum value of integration time which could be used in

autoIntegration mode.


"flickerFrequency"

- allows to set frequency [Hz] of measured light flicker. This frequency is

used in automatic mode to determine the integration time which is a

multiple of the flicked period.


Stage Interpolation is not important.

Section "global" contains information about mounted accessory code (value name="measuringHeads")
and allows to start dark current calibration (value name="Calibration"). Changing text "Select calibration.."
to "Dark Current" will execute dark current calibration on the device.

Example of received config message:



unit="[ms]">10
1
true
max="10000" unit="[ms]">10000
50.000000


unit="[nm]">340
unit="[nm]">750




010.cfg




Select calibration..