773 pH/mV Simulator

10

2.8 Checking the 617 and 679 Rancimats

The measurement of the conductivity and the presentation on the
printer can be checked channel by channel. The conductivity is read
off from the display.
The temperature of the heating block is not important. It may be
necessary to create the starting condition (for 679 T> 50 °C).

773 pH/mV Simulator

x100

x10

x1

-

-

14,3k

PT1000

PT100

RE

WE

WE

ON/OFF

pH

U/mV

-

+

20

6

4

2

18

16

14

12

10

8

0

+

+

0

20

25 40 60

80

100

C

1000

~0,002

pH=7

U=0mV

i

R /M

1. Connect the socket "14.3 k

" of the 773 pH/mV Simulator with

the measuring input of the Rancimat with the 6.2150.010 Cable.

2. Note the cell constant and the printer settings. Set the cell con-

stant to 1 and make the following printer settings:
cond.range

20 uS/cm

paper feed

20 cm/h

Press on the Rancimat.

3. The measured value should be approx. 70 uS (exact value in the

table - field R-Test / G).
Allow each channel to print out 2-3 times.

4. When the printer is not printing out an active channel make the

following connections:

773 pH/mV Simulator

x100

x10

x1

-

-

14,3k

PT1000

PT100

RE

WE

WE

ON/OFF

pH

U/mV

-

+

20

6

4

2

18

16

14

12

10

8

0

+

+

0

20

25 40 60

80

100

C

1000

~0,002

pH=7

U=0mV

i

R /M

The measured value should be approx. 65 uS (exact value in the
table - field Add.data / green).
Allow each channel to print out 2-3 times again.

5. Repeat steps 3 and 4 for all channels.
6. Measure the alteration in conductivity in the Rancimat printout.
7. Reset the actual values for the cell constant and printer settings

on the Rancimat.

6.2150.010 Cable

6.2150.010 Cable