beautypg.com

Technical data, Use according to the intended purpose – Retsch MM 400 User Manual

Page 9

background image

2/8/2012

9

Retsch GmbH

Technical data

Target group: Operating authority, operator

Machine type designation: MM 400

Use according to the intended purpose

The MM 400 is a laboratory machine and is suitable for milling

and homogenising soft, fibrous, hard and brittle materials in

the wet and dry state.
The MM 400 is designed for use with screw-type milling cups

with a steel casing and a milling set with clamping centerings

of dia. 36 mm and a min. clamping length of 60 mm. It is used

for fast, super fine milling of 2 samples simultaneously.

Ceramic milling cups without a complete steel casing must not

be used.
The closed milling system guarantees complete recovery of the

samples. Owing to the extremely short milling time and the

high final fineness of the material to be milled the MM 400 is

also ideally suitable for sample preparation for all spectral

analyses.
Final finenesses of down to 1 µm can be achieved, depending

on the milling time and the specific properties of the sample

material. The optimum milling cup filling is as a rule 1/3 of the

milling cup volume.
Exceptions to this are voluminous materials, such as wool,

leaves, grasses and suchlike. In these cases a filling level of

70-80% is necessary.

The sample quantity should not be less than 25% of the

milling cup volume.
The milling ball/s may otherwise damage the

milling cups.

The milling cup materials and milling ball materials

must always be identical.
The milling ball/s may otherwise damage the

milling cups.

Do not make any alterations to the machine and use

only spare parts and accessories approved by Retsch.
Otherwise the Declaration of Conformity with the

European directives by Retsch loses its validity.
Futhermore this will result in the loss of any kind

of guarantee claim.

Do not clamp any ceramic milling cups of the MM 2/200

or any milling cups without a complete steel casing.
Clamping by the metallic components of the

milling cup holder of the MM 400 can cause

damage to the ceramic parts.