Procedure:

1. Turn OFF instrument.

2. Pry sensor out of its socket using a small flat-blade screw-

driver (Fig 4) and discard.

3. Remove battery cover.

4. Remove old matching resistor (Fig 5) and discard.

5. Locate new sensor’s matching resistor; then cut and bend

the resistor leads using the forms that are molded into the

battery cover (Fig 6). Note that If the resistor leads are cut

too short, the instrument may not function.

6. Insert new matching resistor into printed circuit board,

and plug new sensor into socket at end of probe. Notice that

the sensor’s six pins and the socket are symmetrical and not

keyed, thus permitting the sensor to be plugged in two ways.

The socket, however, is wired so that it doesn’t matter which

way the sensor is installed.

7. Install battery cover.

8. Test instrument response (refer to Section 4.2).

Figure 4. Removing Sensor

Maintenance

10

Leakator 10

Instruction 19-9167

Pry sensor from its socket

using a small flat-blade

screwdriver inserted

through access hole.