Leica Biosystems IPS (Supplement) User Manual
Page 8

8
2013-01 Supplemental sheet for IP C and IP S V 1.1 RevB – 09/2013
Fig. 3
5
Removing the red protective cap
• Unscrew the red protective cap all the way, remove
the information flag and place the red protective cap
in the recess provided on the ink cartridge (6,
Fig. 8
).
• Upon completion, make sure that the red retaining
bracket is in the correct position (8,
Fig. 8
) and close
the door on the printer.
Inserting the new ink cartridge (continued)
Puncturing the cartridge seal requires that
some force be applied. Withdraw the ink car-
tridge about 30 mm from the slot and push it
back in as far as it will go; repeat this step two
or three times (5,
Fig. 3
).
Follow the information on the flag attached to
the ink cartridge (
see Fig. below
).
• Open the red protective cap (4,
Fig. 2
) one turn coun-
terclockwise.
• Then fully insert the ink cartridge into the slot.
Informational flag
Fig. 8
6
8
Supplemental sheet for instructions for use for IP C and IP S V1.9 RevD - 05/2013
• For the procedure,
see Instructions for Use Chap. 6.2
and
Chap. 6.2
.
Fig. 9
Fig. 10
Removing the location plate
• For the procedure (
Fig. 9
),
see Instructions for Use,
Chap. 6.2
.
Manually cleaning the print head
• For the procedure (
Fig. 10
),
see Instructions for Use,
Chap. 6.2
.
Inserting the location plate
Never press the CLEAN button while a new or
used ink cartridge is in the instrument!