Labnet MPS 1000 Mini PCR Plate Spinner User Manual

Page 3

background image

Note: A slight vibration may be felt when the centrifuge reaches full speed. Excessive
vibration indicates that the rotor is not balanced. DO NOT run the certrifuge unbalanced.
Always use a balance plate of the same brand, type and sample volumes.

Cleaning

:

The outside of the centrifuge may be wiped down with a clean, soft cloth. The rotor slots
may be cleaned with a sponge. Do not pour liquids in or over the centrifuge Do not
immerse the centrifuge in liquid.

Safety Precautions:

NEVER

use the centrifuge in any manner not specified in these instructions.

NEVER

put hands in the rotor area unless the rotor is completely stopped.

NEVER

move the centrifuge while the rotor is spinning.

NEVER

use solvents or flammables near this or other electrical equipment.

NEVER

centrifuge flammable, explosive or corrosive materials

NEVER

centrifuge hazardous materials outside of a hood or proper contain

ment

facility
NEVER

continue to operate a centrifuge that is vibrating excessively.

ALWAYS

load the rotor symmetrically. Each plate should be counterbalanced by

another plate of the same brand, type and volume.

ALWAYS

use plates that have been sealed in some manner to avoid spillage.

ALWAYS

load plates so that the sealed side faces the center of the rotor and the

bottom of the plate faces out.

ALWAYS

be sure the plates are pushed all the way to the bottom of the rotor slot

before operating the centrifuge.

ALWAYS

locate the centrifuge within easy access to an electrical outlet.

WARNING!

:


This unit is intended for short run usage only, requiring 15-20 seconds to
spin down samples. Inadvertent continuous use could cause sample
temperature to rise and the thermal switch to open, which will shut down
the motor

Questions and Service:
Should you have a question about the operation of the MPS-1000 Mini Plate Spinner
Centrifuge or if the centrifuge should require service, contact Labnet’s Technical Service
Department at 888-LABNET1. Do not send in a unit for service without first callint to
obtain a repair authorization number. Should the centrifuge require return to Labnet for
service, it should be properly packed to avail damage. Any damage resulting from
improper packaging shall be the responsibility of the user.