Troubleshooting
Abnormality
Reason
Checkout
LCD shows “Er 5”
LCD shows “Er 6”
LCD shows “Er 7”
LCD shows “Er 8”
LCD shows “Er A”
No response when
you press button or
load battery
Cuff pressure above
300mmHg
More than 3 minutes with
cuff pressure above 15 mmHg
EEPROM accessing error.
Device parameter
checking error
Pressure sensor
parameter error
Incorrect operation or
strong electromagnetic
interference
Measure again after 5
minutes. If the monitor
is still abnormal, please
contact the local
distributor or the factory.
Take out batteries for
5 minutes, and then
reinstall all batteries.
17
Harvard Medical Devices Ltd.
Unit 1301-2, 13th Floor, Railway Plaza,
39 Chatham Road South, Tsimshatsui,
Kowloon, Hong Kong.
Kinetik Medical Devices Ltd.
Harvard House, The Waterfront, Elstree Road,
Elstree, Herts WD6 3BS, UK
1. Do not drop this blood pressure monitor or subject it to heavy impact.
2. Avoid exposure to high temperatures, direct sunlight and contact with water.
3. If the monitor is stored at temperatures below freezing, do not use immediately. Leave at room temperature
for at least 1 hour before use.
4. Do not attempt to disassemble the monitor.
5. Do not leave the batteries in the monitor if it is not in use for a long period of time.
6. It is recommended the performance should be checked every 2 years or after repair.
7. If the monitor becomes dirty, please clean it with a soft dry cloth. Do not use any abrasive or volatile cleaners.
8. The supplier will make available on request the circuit diagrams, component part lists, descriptions, calibration
instructions, or other information which will assist the USER’s appropriately qualified technical personnel to
repair those parts of EQUIPMENT which are designated by the manufacturer as repairable.
18
Maintenance