JULABO MA-26 Heating Circulators User Manual
Page 28
Safety installations, warning functions
28
8.2.
Switch-over from warning to shutdown function
If a shutdown of functional elements (e.g. heater, circulating pump) is
required when the limit values are exceeded or undercut the
circulator can be changed over from warning function >WARNING<
to shutdown function >ALARM<.
Factory setting:
>0 = WARNING<
1. Press the key
.
2. Select the menu >Aty (ALARM-TYPE)< by pressing the
key.
3. Press the
key and the set parameter will flash
.
(Example: 0)
4. Change the parameter by pressing the
key and confirm by
pressing the
key.
or
press the
key if the parameter is to retained.
Setting >0 = WARNING<
A mere warning function with optical and audible warning signal
(interval tone) A message appears on the MULTI-DISPLAY
(LED):
or
OVERTMP
SUBTEMP
Setting >1 = ALARM<
Temperature limit with shutdown of heater and circulating pump.
An audible alarm sounds (continuous tone) and a message
appears on the MULTI-DISPLAY (LED):
or
OVERTMP
SUBTEMP