QS5 Operating Manual

The exhaust gas from the vacuum pump is composed of oil and gases

from within the pump. The exhaust gas will be hot and should be well
ventilated.

Never operate the pump with a blocked or restricted exhaust. The

resulting back pressure may blow out the sight glass or otherwise damage the
pump.

The QS5 has been designed to operate most efficiently on high viscosity

vacuum pump oil. Other vacuum pump oil may work, but performance results
may not be as stated.

The QS5 is not designed for pumping corrosive, aggressive or explosive

gases, or for use in flammable environments. Do not use with ammonia or
lithium bromide systems.

4.0 Identification Of Parts

Oil fill cap

Exhaust demister

Oil fill level

Oil sight glass

Oil drain

Electrical housing (contains power
switch, power socket and power
supply conversion switch)

Not shown:

Gas ballast

Electrical Socket

110V/220V switch

Inlets (1/4 in and 3/8 in)