Alarm Mode

1) Press ON/OFF key to power up the MMX-6.

2) Press the ALRM key. ALAr OFF, or ALAr followed by a thickness

value and flashing IN (or MM) symbol will be displayed - indicating

the alarm mode is enabled.

3) Repeat step 2 to toggle between ALAr OFF or ALAr value and

flashing IN / MM symbol (On).

4) Assuming ALAr value flashing IN is displayed, use the UP and

DOWN arrow keys to scroll to the desired nominal alarm value.

5) Press the ALRM key once again to select the nominal value

entered.

6) The MMX-6 is now ready to perform measurements using the

Alarm feature.

Dual-Multi Mode

Often times users and inspectors in the field are faced with coated

materials such as pipes and tanks. Typically inspectors will need to

remove the paint or coating prior to measuring, or allow for some fixed

amount of error introduced by the paint or coating thickness and velocity.

An A-Scan scope with a special echo -echo mode is generally used in order

to accurately perform measurements through paints and coatings.