2050 Operations Manual

Rev I 6/3/2014

Page 17 of 21

SECTION 7

– PREVENTATIVE MAINTENANCE


The CARON circulator has been robustly designed to minimize performance problems.
However, regular maintenance is very important for continuous trouble free operation.


Recommended Daily Maintenance Checks

Check the Temperature display versus setpoints.
Check for and correct any alarm condition.



Recommended Monthly Maintenance Checks

Check condenser air intake filter on top of unit. If it is dirty, replace it with CARON

part number FLTR306. Washing the filter will result in poor performance.


Recommended Annual Maintenance Checks

Disinfect all interior surfaces with a general purpose laboratory cleaning agent such

as Cole-Parmer #G-08796-00.

Vacuum or blow out fan and internal components.
Perform a temperature calibration

R-134a refrigerant

This unit contains R-134a which is a fluorinated greenhouse gas covered by the

Kyoto Protocol (Regulation EC No 842/2006) with a GWP of 1,300. Quantity of R-
134a is less than 0.2 kg.

To prevent electrical shock, turn off and unplug the unit before performing any service
or maintenance work on the unit. Performing operations or maintenance on this unit
other than the ones described in this manual can cause hazard situations and should
only be performed by qualified service personnel.