D-2. error codes – C.B.S. Scientific EPS - 600 User Manual
Page 11

EPS-600
9
D-2. Error codes
The EPS-600power supply has built-in error messages to help troubleshoot problems with the 
electrical circuit. The error codes and possible corrective actions are listed below:
Error
Code
Explanation
Possible Solution
E1
No load detected
The voltage leads to the 
electrophoresis cell are not 
plugged. 
The current load for the 
separation is below 4 mA the 
lower limit of the power supply.
1. Check all the connections 
2. Verify that buffer volumes are sufficient to 
contact all electrode wires and that buffer 
is contacting both surfaces of the gel. 
3. Press the any button* to clear the error
code.
E3
Change in Load Resistance
Open circuit was detected at some 
point after a successful start. 
Buffer is leaking from the cells. 
There is a loose electrical 
connection in the circuit. 
1. Check and correct all potential resistance
problems.
2. Check buffer volumes. 
3. Check all voltage leads. 
4. Check the power supply adapters fitting 
(If used)
5. Press any button* to clear the error code.
E4
Power Failure during a 
Continuous run (no timer used) 
1. Restart the run, if desired. Press any
button* to clear the error code.
E5
Power Failure during a timed run 
with PFD option disabled. 
Run was not complete. 
1. Restart the run, if desired. 
2. Press any button* to clear the error 
code.
See page 7 for information on the PFD option.
E6
Power Failure during a timed run 
with PFD option enabled. 
Despite the power outage the run 
was completed successfully. 
1. Press any button* to clear the error
code.
See page 7 for information on the PFD option.