3 weight / dimensions, 4 application parameters, 1 centrifuge 5424 – Eppendorf 5424 R Centrifuge User Manual
Page 35: Weight / dimensions, Application parameters 9.4.1, Centrifuge 5424, 9 technical data, Operating manual
35
9 Technical data
Centrifuge 5424 / 5424 R — Operating manual
9.3
Weight / dimensions
Centrifuge 5424
Centrifuge 5424 R
*) The noise level was measured according to DIN EN ISO 3745 frontally in a sound measuring
room with accuracy class 1 at a distance of 1 m from the device and at lab bench height.
9.4
Application parameters
Acceleration and deceleration times of Centrifuge 5424 / 5424 R with the rotor FA-45-24-11
(according to DIN 58 970)
These values were calculated for 230 V at 23°C.
9.4.1
Centrifuge 5424
Dimensions:
Width: 236 mm (9.30 in.)
Depth: 320 mm (12.60 in.)
Height: 227 mm (8.94 in.)
Weight without rotor:
13.4 kg (29.5 lb)
Noise level:
< 51 dB(A) *
Dimensions:
Width: 290 mm (11.42 in.)
Depth: 480 mm (18.90 in.)
Height: 260 mm (10.24 in.)
Weight without rotor:
21.0 kg (46.3 lb)
Noise level:
< 54 dB(A) *
Rotor
Without soft ramp
With soft ramp SOFT
FA-45-24-11
15 s / 16 s
5424 : 24 s / 28 s
5424 R: 26 s / 28 s
Run time:
30 s to 9:59 h, infinitely (oo),
adjustable up to 10 min. in 0.5 min.
increments, thereafter in 1 min. increments.
Relative centrifugal force (RZB or rcf):
1 to 20,238 x g (230 V, 100 V)
1 to 21,130 x g (120 V)
adjustable in 50 x g increments.
Speed:
100 to 14,680 rpm (230 V, 100 V)
100 to 15,000 rpm (120 V)
adjustable in 50 rpm increments.
Max. load:
24 micro test tubes of 2.0 mL each.
Max. kinetic energy:
4070 Nm
Test log mandatory:
No
Permitted density of the material for
centrifuging (at max. g-force (rcf) or
speed (rpm) and max. load):
1.2 g/mL
EN
O
perating manual