9 technical data, 1 power supply, Technical data 9.1 – Eppendorf 5418 R Centrifuge User Manual
Page 37: Power supply, 9technical data 9.1 power supply
Technical data
Centrifuge 5418 / 5418 R
English (EN)
37
9
Technical data
9.1
Power supply
Centrifuge 5418
Centrifuge 5418 R
Mains connection
230
V,
50
Hz – 60
Hz
120
V,
50
Hz – 60
Hz
100
V,
50
Hz – 60
Hz
Current consumption
1.4
A
(230
V
)
2.8
A
(120
V
)
3.4
A
(100
V
)
Power consumption
Maximum 170
W
EMC: Interference emission. (radio interference)
EN 61326 – Class B (230
V)
FCC15 - Class A (120
V/100
V)
EMC: Noise immunity
EN 61326
Overvoltage category
II
Fuses
2.5 AT (230 V)
5.0 AT (120 V / 100 V)
Degree of pollution
2
Mains connection
230
V,
50
Hz – 60
Hz
120
V,
50
Hz – 60
Hz
100
V,
50
Hz – 60
Hz
Current consumption
1.4
A
(230
V)
2.8
A
(120
V)
3.0
A
(100
V)
Power consumption
Maximum 320
W
EMC: Interference emission. (radio interference)
EN 61326 – Class B (230
V)
FCC15 - Class A (120
V/100
V)
EMC: Noise immunity
EN 61326
Overvoltage category
II
Fuses
2.5 AT (230 V)
6.25 AT (120 V / 100 V)
Degree of pollution
2