The following rotors and accessory buckets have a maximum service life of 7 years:

A-4-38

5702 720.003 and 5702 740.004

A-8-17

5702 700.002

The date of manufacture is engraved on them in the format 10/01 (October 2001).

Transparent aerosol-tight caps made of polycarbonate have a maximum service life of 3 years.
The date of manufacture is embossed on them in the form of a clock

.

Aerosol tight caps may lose their sealing strength when exposed to organic solvents (e.g. phenol, chloroform).
Check the aerosol-tight caps regularly for changes or cracking. Aerosol-tight caps showing cracks or other changes must
be replaced immediately.

Rotors, caps or buckets which have been damaged by chemical or mechanical factors or which have passed

their maximum service life may no longer be used

!

Transfer

If the device is passed on to someone else, please include the instruction manual.

Disposal

In case the product is to be disposed of, the relevant legal regulations are to be observed.

Information on the disposal of electrical and electronic devices in the European Community

The disposal of electrical devices is regulated within the European Community by national regulations based on EU
Directive 2002/96/EC on waste electrical and electronic equipment (WEEE).
According to these regulations, any devices supplied after 13.08.05 in the business-to-business sphere, to which this
product is assigned, may no longer be disposed of in municipal or domestic waste. They are marked with the
following symbol to indicate this.

As disposal regulations within the EU may vary from country to country, please contact your supplier if

necessary.

Declaration concerning the ATEX directive (94/9/EC)

The present design and ambient conditions inside Eppendorf centrifuges mean that they are not suitable for use in a
potentially explosive atmosphere. The centrifuges must therefore only be used in a safe environment such as the open
environment of a ventilated laboratory or a fully-extracted fume hood. The use of substances which may contribute to a
potentially explosive atmosphere is not permitted. The final decision on the risks in connection with the use of such
substances is the responsibility of the user of the centrifuge.