6
Current certificates can be found at www.eppendorf.com. Technical specifications subject to
change and errors excepted!
1
SBS: Society for Biomolecular Screening.
2
The centrifugation stability of any plate type is generally dependent on the plate itself,
centrifuge, rotor, rotor suspension gear, adapter, maximum rotation / maximum gravitation field,
operating temperature, centrifugation time, density of centrifuged liquid, organic solvent,
accumulated total weight and correct operation.
3
Centrifuge stacked plates at lower speed only.
3
Product description
Instructions for use
The Eppendorf Deepwell Plates and Microplates are uncoated polypropylene disposables for
analyzing, preparing, mixing, centrifuging, transporting and storing of solid and liquid probes.
Eppendorf Microplates can be used in all conventional plate readers. Only trained technical staff
may use these consumables.
The Eppendorf Deepwell Plates and Microplates are characterized by a high chemical and
temperature resistance and can be sealed using suitable sealing options. The consumables fulfill
the ANSI/SBS standards for microplates.
This product can be used in development, research, routine and training laboratories in industry,
in chemistry, and in training and instruction. The applications extend over the areas of life
science, biotechnology, chemistry and clinical research, but are not limited to these areas.
Material
Polypropylene, colored and colorless
Chemical resistance
The plates, including border, show a high resistance to UV
light and chemicals. Compare Application No. 56: The best
material for original Eppis! on www.eppendorf.com. In case
of doubt, contact Eppendorf Application Support.
Dimensions
In accordance with ANSI/SBS standard
1
I to IV, see
illustrations.
Centrifugability
at 4 °C to 40 °C
2
Eppendorf Deepwell Plates standard, DNA LoBind and
Protein LoBind quality: to 6000 x g.
Eppendorf Deepwell Plates sterile: to 5000 x g.
Eppendorf Microplates PCR clean, sterile, DNA LoBind and
Protein LoBind quality: to 6000 x g.
3
Autoclaving
Autoclavable (121°C, 20 min.). The dimensional stability of
the disposables can be affected.
Operational temperature
-86°C to +100°C
Storage before use
Store dry at room temperature. Protect from direct sunlight
and UV light.
Eppendorf Standard Quality
Consumables in original Eppendorf Quality for routine
application.
Eppendorf PCR clean
(lot certified)
Free of RNase, DNase, human DNA and PCR inhibitors.
Eppendorf Protein LoBind and
DNA LoBind
Consumables with the highest recovery rates for proteins
and/or DNA/RNA molecules. They are PCR clean, i.e. free
from RNase, DNase, human DNA and PCR inhibitors.
Eppendorf sterile (lot certified) Consumables are sterilized by irradiation. Best before: see