Atex – Liebherr GGU 1550 Premium User Manual
Page 24

46
47
DIN
Atex
DIN stands for ‘Deutsches Institut für Normung’ (German Institute
for Standardization). The Institute’s specialist bodies define industry
standards. DIN 58345 relates to pharmaceutical refrigerators that
are used to store pharmaceuticals at an operating temperature of 2 °C to 8 °C. DIN
58345 currently reflects the latest technological advances and establishes precisely
formulated requirements in relation to pharmaceutical refrigerators.
Basic requirements established by DIN 58345:
✓
Operating temperature: +2 °C to +8 °C
✓
Usable in ambient temperatures of +10 °C to +35 °C
✓
Noise emissions: < 60 dB(A))
✓
Visual and acoustic warning system for high and low temperature alarm (min/
max thermometer)
✓
Warning system for min. 12h in the event of a power cut
(power cut alarm device)
✓
Safety thermometer for preventing freezing temperatures
✓
Component mechanical load-bearing capacity
✓
Lockable doorr
The MKv and MKUv models comply with DIN 58345. Conversely, all other Liebherr
laboratory refrigerators and freezers do not comply with DIN 58345.
Several models can be installed beneath
a worktop. In this case, a ventilation grill
with cross-section of at least 200 cm
2
is
required in the worktop for ventilation at
the rear of the appliance. If no ventilation
grill is going to be fitted in the worktop,
the recess height must be at least 860
mm to ensure sufficient heat dissipation
to the front. Under-worktop laboratory ap-
pliances offer a space-saving solution,
which is particularly suitable for use in
restricted spaces.
The appliances are factory-fitted with
door hinges on the right. Being able to
change the hinge side allows the appli-
ance to be positioned even more optimally
and tailored to individual requirements.
Pharmaceutical storage: DIN 58345
Under-worktop models
Interchangeable door hinges
The Atex 95 certification certifies that Liebherr laboratory refrigerators
and freezers with explosion protection as defined by EU Directive 94/9/
EC (ATEX 95) are suitable for the storage of explosive and highly flam-
mable materials.
Most important precautions:
✓
Interior free from ignition sources
✓
No openings of the interior area to the exterior
✓
Separate condensate dish for emptying the condensate in the interior
✓
Interior components made from glass, plastic or with plastic coating
✓
Metal parts with protective earth connection
➞
No risk of sparks being formed against metal
Explosion protection: ATEX 95
Laborator
y
- WTes 5872 Vinidor WKt 6451 GrandCru FKDv 3713 Premium GG 4010 WTr 4211 Vinothek WTb 4212 Vinothek GGv 5860 BCDv 4313 FT 2900 WKr 3211 Vinothek GGPv 1470 ProfiLine GGv 5010 WKb 3212 Vinothek MKv 3910 GGv 5810 ZKes 453 Humidor FT 2902 G 5216 WTes 5972 Vinidor WKb 1812 Vinothek WTpes 5972 Vinidor Fv 3613 WKb 4212 Vinothek WKt 5551 GrandCru MKUv 1610 WKes 4552 GrandCru WKes 653 Grand Cru FDv 4613 WKt 4552 GrandCru GG 4060 WKr 4211 Vinothek WKt 5552 GrandCru GG 5260 GG 5210 WKt 4551 GrandCru GGPv 1490 ProfiPremiumline GGU 1500 Premium BCDv 1003 GGPv 6570 ProfiLine WKr 1811 Vinothek GGPv 6590 ProfiPremiumline GGv 5060