Operator’s Manual for the Pressure Injection Cell

BE SAFE. ALWAYS MAKE SURE THAT THERE IS NO PRESSURE IN THE CELL BEFORE
LOOSENING THE BOLTS, AND THAT THE BOLTS ARE SECURELY FASTENED BEFORE

APPLYING PRESSURE TO THE CELL. USERS MUST ALWAYS WEAR PROTECTIVE

APPAREL, INCLUDING EYEWEAR AND MUST BE TRAINED IN WORKING WITH HIGH

PRESSURE.

TROUBLESHOOTING

In addition to the tips given below,

a thorough list of troubleshooting tips is
at

http://www.nextadvance.com/api/index.
cfm/cms.page/i/5063/pressure-
injection-cell-troubleshooting.

If gas leaks through the top-base

interface of the pressure injection
cell,
the three bolts might not be fully
tightened or the O-ring might be nicked
or dirty or missing. If the O-ring is
dirty, wash it off with a soap and water
and return it to its groove. Contact us
for a replacement O-ring.

If gas leaks around the capillary,

the ferrule should be replaced.

If you do not see sample traveling

through the capillary after a few
minutes,
check that the gas tank is
supplying the

desired pressure

(typically on the order of several
hundred psi) and that the 3-way valve
knob is pointing toward the pressurized
gas line.

SUPPORT

A thorough list of FAQs is at

http://www.nextadvance.com/api/index.
cfm/products.info/c/427/Pressure-
Injection-Cell-Capillary-Loader#/FAQ.

If you cannot find a good answer

there, please contact customer service

by email at

[email protected]

or by telephone.

SPECIFICATIONS

Max. Pressure: 6.8 MPa (1000 psi)

Size:

6.5 in (16 cm) long x 3 in (8

cm) wide x 4 in. (10 cm) high

Weight:

2-1/2 lbs. (1 kg)

Relative Humidity: 5 - 90% non-
condensing

Operating Temperature: 4 - 40°C

Altitude: <2000m

Storage Temperature: -40 to 50°C

Meets

requirements

Power Requirement (PC1000-MAG
model): 24V DC, 0.8 A

WARRANTY

Next Advance warrants its Products

against defects in materials and
workmanship for time periods which
vary according to the Product. Within
these time periods, Next Advance will
replace or repair, without charge to the
original purchaser, any part which is
defective.
Pressure Injection Cells Two years

The warranty is void if the Product

is defective due to product accident,
product modification, exposure to
radiation other than for sterilization,