Operator’s Manual for the Pressure Injection Cell
BE SAFE. ALWAYS MAKE SURE THAT THERE IS NO PRESSURE IN THE CELL BEFORE
LOOSENING THE BOLTS, AND THAT THE BOLTS ARE SECURELY FASTENED BEFORE
APPLYING PRESSURE TO THE CELL. USERS MUST ALWAYS WEAR PROTECTIVE
APPAREL, INCLUDING EYEWEAR AND MUST BE TRAINED IN WORKING WITH HIGH
PRESSURE.
TROUBLESHOOTING
In addition to the tips given below,
a thorough list of troubleshooting tips is
at
http://www.nextadvance.com/api/index.
cfm/cms.page/i/5063/pressure-
injection-cell-troubleshooting.
If gas leaks through the top-base
interface of the pressure injection
cell, the three bolts might not be fully
tightened or the O-ring might be nicked
or dirty or missing. If the O-ring is
dirty, wash it off with a soap and water
and return it to its groove. Contact us
for a replacement O-ring.
If gas leaks around the capillary,
the ferrule should be replaced.
If you do not see sample traveling
through the capillary after a few
minutes, check that the gas tank is
supplying the
desired pressure
(typically on the order of several
hundred psi) and that the 3-way valve
knob is pointing toward the pressurized
gas line.
SUPPORT
A thorough list of FAQs is at
http://www.nextadvance.com/api/index.
cfm/products.info/c/427/Pressure-
Injection-Cell-Capillary-Loader#/FAQ.
If you cannot find a good answer
there, please contact customer service
by email at
or by telephone.
SPECIFICATIONS
Max. Pressure: 6.8 MPa (1000 psi)
Size:
6.5 in (16 cm) long x 3 in (8
cm) wide x 4 in. (10 cm) high
Weight:
2-1/2 lbs. (1 kg)
Relative Humidity: 5 - 90% non-
condensing
Operating Temperature: 4 - 40°C
Altitude: <2000m
Storage Temperature: -40 to 50°C
Meets
requirements
Power Requirement (PC1000-MAG
model): 24V DC, 0.8 A
WARRANTY
Next Advance warrants its Products
against defects in materials and
workmanship for time periods which
vary according to the Product. Within
these time periods, Next Advance will
replace or repair, without charge to the
original purchaser, any part which is
defective.
Pressure Injection Cells Two years
The warranty is void if the Product
is defective due to product accident,
product modification, exposure to
radiation other than for sterilization,
connection to an improper electrical