beautypg.com

Aspiration heads, Replacement, Sanitizing – Next Advance DX User Manual

Page 11

background image

O

PERATOR

S

M

ANUAL FOR THE

F

REEDOM

R

OCKER

DX

The quick release pins must be firmly

pushed through the holes until the rings
are against the metal housing.

Route the tubing past the cams as
shown in the photo below. Use the
Freedom Rocker in manual mode to
rotate the cams by setting the fill
volume to 0 mL and running FILL
with the white, blue, and yellow
sources.

Then, connect the luer lock fittings
to the aspiration head and the
others ends onto the One>Click™
caps on the bottles.

ASPIRATION HEADS

The aspiration head can be swung
out of the way when accessing the
trays.

The quick release pin (white arrow)
holds the aspiration head to the hinge.

Replacement
To switch aspiration/dispense heads,
first disconnect the tubing from the
head by undoing the luer lock
fittings. Then simply remove the
quick release pin by pulling on the
ring. Next, place the desired

aspiration head on the platform, line
up its transverse hole with that in
the pivot block, and replace the
quick release pin. Finally, reconnect
the luer lock fittings.

Sanitizing
To sanitize the tube set or the
aspiration head, use a standard
sanitizing solution, e.g. 5% bleach
or 30% ethanol, followed by
thorough rinsing with distilled water.
Please note that the tube set and
the aspiration head should not be
autoclaved because the temperature
is too high for some of the plastic
parts.

11