beautypg.com

Operator’s manual for the bullet blender – Next Advance BBY User Manual

Page 5

background image

Operator’s Manual for the Bullet Blender

®

NOTES:

The Bullet Blender Storm 5 is

designed to work with 5 mL Axygen

®

brand tubes only. At high speed
settings, there may be some flaking of
the tubes. This is a normal side effect of
homogenization. The higher speed
enables homogenization of tougher
tissue.

CLEANING

If you wish to clean your Bullet

Blender, clean the outside of the unit
only with mild soap, water and a soft
cloth. Under normal conditions, the
Bullet Blender should never need to be
disassembled for cleaning. In the case
of a large spill, unplug the instrument,
remove the sample tube plate with a
1/8” hex wrench, wipe out the spill
using standard laboratory safety
precautions, and replace the sample
tube plate. Do not touch or tamper with
the electronics.
Contact tech support for a detailed
protocol:

[email protected]

TROUBLESHOOTING

In addition to the tips given below,

thorough list of troubleshooting tips is
at http://bbtroubleshooting.nextadvance
.com
If the Bullet Blender doesn’t start, the
plug of the power supply cord may not
be in a live wall outlet or the power
supply connector may not be fully
inserted in the socket on the back of the
Bullet Blender.
If the unit stops working,
turn the
system off for 15 minutes to allow the
electronics to reset. If the Bullet

Blender does not turn on after this

period, contact

customer service.

If the caps on snap-cap tubes pop
open or the caps on screw-cap t

ubes

loosen, make sure that the interface
regions or screw threads between the
lids and the caps is dry when you close
the caps or screw them on, so that there
is enough friction for the caps to remain
tight. Using recommended types of
tubes will minimize cap failure.

SUPPORT

FAQs, protocols, and other helpful

information are available on our
website, http://www.nextadvance.com.
Click on the Bullet Blender, then on the
appropriate link. If you cannot find an
answer there, please contact customer
service by email at support
@nextadvance.com or by telephone at
1.518.674.3510 or (800) 738-1681.

SPECIFICATIONS

Size: 28 cm (11 in.) deep x 23 cm (9
in.) wide x 21.5 cm (8.5 in.) high.
Weight: 4 to 5 kg (8.5 to 11 lbs.),
depending upon model.
Power Requirement: BBX24 and
BBX24B models: 24 VDC, 0.75 Amp
BBY24M BBY5E and BBY5M
models: 24 VDC, 2.5 Amp
Capacity: 24 of 1.5 - 2.0 mL poly-
propylene high quality (e.g.

Eppendorf

®

) snap-cap tubes, 24 of 1.5

mL RINO

®

screwcap tubes, 12 of 5 mL

Eppendorf

®

tubes or 12 of 5 mL

Axygen

®

5 mL transport tubes,

depending upon model.

5