Features:
1. A simple turn of a knob adjusts power.
2. Lighted
ON/OFF
Switch.
3. Circuit
protection.
4. 10-Amp output. 1200 Watts Maximum
5. Low profile housing.
General Use:
The Power Controller is used to manually adjust power to a load with a simple turn of a knob. It was designed for use with heat-
ers / heating mantles as well as other resistive loads. It outputs 5% to 100% of the rated voltage. The approximate values for the
voltage output can be found in Table 1¹.
Table 1: Approximate Output Voltages
¹
Actual voltage outputs can be found by using a voltmeter or digital multi-meter across the heating mantle or load.
Knob Setting (%)
Approximate VAC
Output
0
10
+/-
2
+/-
5%
10
14
+/-
2
+/-
5%
20
21
+/-
2
+/-
5%
30
30
+/-
2
+/-
5%
40
42
+/-
2
+/-
5%
50
51
+/-
2
+/-
5%
60
64
+/-
2
+/-
5%
70
80
+/-
2
+/-
5%
80
97
+/-
2
+/-
5%
90
112
+/-
2
+/-
5%
100
115
+/-
2
+/-
5%