Troubleshooting – BrandTech Scientific accu-jetpro User Manual
Page 46

Troubleshooting
Trouble
Possible cause
Action to be taken
Suction capacity
impaired
Motor speed is reduced Increase motor speed (p. 35)
Check valve closed
Set mode selection switch
for 'power delivery' and press
lower button for blow out
(p. 37)
Filter or check valve
dirty
Clean and/or replace check
valve (p. 41)
Pipette drips
Filter not properly
positioned
Insert filter properly (p. 42)
Adapter or pipette
damaged
Replace adapter or pipette
(p. 41)
Pipette not held
properly
Adapter dirty or dam-
aged
Clean the adapter, allow it to
dry and/or replace it (p. 41)
After connecting
the AC adapter,
the LED-display
of the instrument
flashes or does not
light up
Plug of the battery not
firmly pushed in
Firmly push in the battery plug
(p. 33)
Wrong AC adapter
Only use original AC adapter
(p. 45)
Battery/AC adapter
defective
Replace battery/AC adapter
Instrument defective
Send the instrument in for
repair (p. 47)
Low battery charge Battery problem
Replace battery (p. 40)
Instrument doesn't
work
Internal error
(carry out reset)
Pull out battery plug and press
a pipetting button for 10 sec-
onds. Then firmly push in the
battery plug again (p. 33)
If recommendations in the Troubleshooting section do not solve current
problems, contact your authorized dealer or the manufacturer directly.