Maintenance, Technical data – A&D A & D Digital Blood Pressure Monitor UA-767 Plus BT User Manual

Page 14

background image

English

12

Maintenance

Do not open the device. It uses delicate electrical components and an
intricate air unit that could be damaged. If you cannot fix the problem
using the troubleshooting instructions, request service from your dealer
or from the A&D service group. The A&D service group will provide
technical information, spare parts and units to authorized dealers.

The device was designed and manufactured for a long service life.
However it is generally recommended to have the monitor inspected
every 2 years, to ensure proper functioning and accuracy. Please contact
either your authorized dealer or A&D for maintenance.




Technical Data

Type

UA-767PBT / UA-767PBT-G

Measurement method

Oscillometric measurement

Measurement range

Pressure: 20 - 280 mmHg

Pulse:

40 - 200 beats / minute

Measurement accuracy

Pressure: ±3 mmHg or 2%, whichever is greater

Pulse:

±5%

Power supply

4 x 1.5V batteries (R6P or AA)

Upper arm circumference 22 - 32 cm (8.7" - 12.6")
Classification Type

BF

Clinical test

According to ANSI / AAMI SP-10 1987

EMC

IEC 60601-1-2: 2001

Wireless communication WML-40AH (MITSUMI Electronics Co. Ltd.)
Operating condition

+10°C to +40°C / 30%RH to 85 %RH

Storage condition

-10°C to +60°C / 30%RH to 85 %RH

Dimensions

Approx. 147 [W] x 64 [H] x 110 [D] mm

Weight

Approx. 300 g, excepting batteries

Note: Specifications are subject to change for improvement without prior notice.

The inspection for every two years of service is only recommended for European
customers, in order to comply with regulatory requirements. Routine inspection
and maintenance is not needed for non-European customers.