Maintenance, Technical data – A&D A & D Digital Blood Pressure Monitor UA-767 Plus BT User Manual
Page 14
English
12
Maintenance
Do not open the device. It uses delicate electrical components and an
intricate air unit that could be damaged. If you cannot fix the problem
using the troubleshooting instructions, request service from your dealer
or from the A&D service group. The A&D service group will provide
technical information, spare parts and units to authorized dealers.
The device was designed and manufactured for a long service life.
However it is generally recommended to have the monitor inspected
every 2 years, to ensure proper functioning and accuracy. Please contact
either your authorized dealer or A&D for maintenance.
Technical Data
Type
UA-767PBT / UA-767PBT-G
Measurement method
Oscillometric measurement
Measurement range
Pressure: 20 - 280 mmHg
Pulse:
40 - 200 beats / minute
Measurement accuracy
Pressure: ±3 mmHg or 2%, whichever is greater
Pulse:
±5%
Power supply
4 x 1.5V batteries (R6P or AA)
Upper arm circumference 22 - 32 cm (8.7" - 12.6")
Classification Type
BF
Clinical test
According to ANSI / AAMI SP-10 1987
EMC
IEC 60601-1-2: 2001
Wireless communication WML-40AH (MITSUMI Electronics Co. Ltd.)
Operating condition
+10°C to +40°C / 30%RH to 85 %RH
Storage condition
-10°C to +60°C / 30%RH to 85 %RH
Dimensions
Approx. 147 [W] x 64 [H] x 110 [D] mm
Weight
Approx. 300 g, excepting batteries
Note: Specifications are subject to change for improvement without prior notice.
The inspection for every two years of service is only recommended for European
customers, in order to comply with regulatory requirements. Routine inspection
and maintenance is not needed for non-European customers.