Introduction, Features – Myron L PT2 User Manual

Page 2

background image

1

INTRODUCTION

Thank you for purchasing the Myron L ULTRAPEN™ PT2 pH

Pen. This instrument is designed to be extremely accurate,

fast and simple to use in diverse water quality applications.

Advanced features include automatic temperature

compensation; highly stable microprocessor-based circuitry;

user-intuitive design; and waterproof housing. A true one-

handed instrument, the PT2 is easy to calibrate and easy to

use. To take a measurement, you simply push a button then

dip the pen in solution. Results display in seconds.

FEATURES

1. Push Button — turns instrument on; selects mode and unit preferences.

2. Battery Cap — provides access to battery for replacement.

3. Pocket Clip — holds pen to shirt pocket for secure storage.

4. Battery Indicator — indicates life remaining in battery.

5. Display — displays measurements, menu options, battery indicator, and

firmware revision (during power-up).

6. LED Indicator Light — indicates when to dip instrument in solution, when

measurement is in progress, and when to remove instrument from solution.

7. pH Sensor — measures hydrogen ion concentration of solution.

8. Soaker Cap — contains pH Sensor Storage Solution to maintain sensor hydration.

To remove, twist the soaker cap while pulling off using caution not to spill the Storage

Solution. To replace, fill the soaker cap half full with storage solution. Twist the soaker

cap while pushing back on using caution as excess Storage Solution may squirt out.

CAUTION: Fill the soaker cap ONLY half full of solution for storage.

CAUTION: Do NOT push the soaker cap beyond the Cap Stop as sensor damage

WILL occur.

NOTE: The formation of KCl crystals around the soaker cap is normal. These

crystals do not affect the sensor life, performance, or accuracy provided they are

rinsed off with water prior to a test.

9. Scoop — used to hold sample solution when dipping is not

possible. To install, push the scoop onto the sensor while

shifting side-to-side. To remove, pull the

scoop off while shifting side-to-side. Verify

the pH sensor remained fully inserted into the PT2. If not, reinstall per pH Sensor

Replacement, p. 5. To use, pour solution into the scoop or hold the scoop directly

under a vertical stream to collect sample.

10. Holster — run your belt through the strap in the back of the holster for hands-

free portability.

11. Lanyard — attach through hole in top of pocket clip.